Nova II – 5 1/2 Digit DC Powered Digital Panel Meter

NOVA II digital panel meter

Features

* 5 1/2 digit display.

 

Specifications

Input

Full Scale Input Voltage : +/- 1.99999 vdc
Input Resistance: >1000 Megohms
Input Bias Current: 20 na (typ)
Input Type: Fully Differential
Input Bias Current: 20 na (typ)
Max Normal Mode Voltage: +/- 150 v
Max Common Mode Voltage: +/- 5 v

Readout

Full Scale Readout: +/- 199,999
Accuracy (Full Scale): Adjustable
Accuracy Stability: +/- 0.005% R +/- 1 digit ( For 3 months @
25 +/- 5 deg C )
Zero (Stability): within +/- 2 digits ( For 3 months @ 25
+/- 5 deg C )
Non Linearity: +/- 0.0001% R +/-
Full Scale Tempco: +/- 10ppm / degree C
Zero Tempco: +/- 1 ppm / deg C
Display type ( digits ): 0.56 ” Hi Eff Red LED
Display Type ( sign ): Minus sign displayed for negative inputs.
No sign displayed for positive.
Decimal Points: Selectable via rear panel connector.
Conversion Rate: 2.5 conv / sec.

Output

All BCD output signals are buffered, latched, fully TTL
compatible signals capable of sinking and sourcing 2 full TTL loads.
The SIGN (POL) output is logical “1” for positive readings and
logical “0” for negative.

Environment / Power

Operating Temp: 0 to +55 deg C
Storage Temp: -40 to +85 deg C
Input Voltage (Power): +/- 15 or +/- 12 vdc @ 35 ma and +5 vdc @
250 ma.

download manual for Nova II